Rat VWF (Von Willebrand Factor) CLIA Kit for Quantitative Detection

SPECIFICATIONS
STJC0001944
🚚 Free UK Delivery on orders over £150
Processing The item has been added
Enquire For Bulk Order
✓ 1-Year Performance Guarantee | SDS & Datasheet Included | Expert Tech Support | Covered by St John's Laboratory Guarantee

General Information

Short DescriptionThis rat VWF kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes in serum, plasma and other biological fluids.
ApplicationsCLIA
ReactivityRat
Sensitivity18.75pg/mL
Detection Limit31.25~2000pg/mL
NoteFOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS.

Product Properties

Storage InstructionIf unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual.
Assay Time3.5h
DetectionChemiluminescence

Target Information

Gene SymbolVwfUniprot ID=VWF_RAT"
SpecificityThis kit recognizes Rat VWF in samples. No significant cross-reactivity or interference between Rat VWF and analogues was observed.
Sample TypeSerum, plasma and other biological fluids

Additional Info

Tissue Specificity Plasma.
Post Translational Modifications N- and O-glycosylated.
Function Important in the maintenance of hemostasis, it promotes adhesion of platelets to the sites of vascular injury by forming a molecular bridge between sub-endothelial collagen matrix and platelet-surface receptor complex GPIb-IX-V. Also acts as a chaperone for coagulation factor VIII, delivering it to the site of injury, stabilizing its heterodimeric structure and protecting it from premature clearance from plasma.
Protein Name Von Willebrand Factor
Vwf
Cellular Localisation Secreted
Extracellular Space
Extracellular Matrix
Localized To Storage Granules
Alternative CLIA Names Von Willebrand Factor CLIA kit
Vwf CLIA kit
Vwf CLIA kit
Specificity This kit recognizes Rat VWF in samples. No significant cross-reactivity or interference between Rat VWF and analogues was observed.
Reproducibility Both intra-CV and inter-CV are

Information sourced from Uniprot.org


Item Specifications Storage
Micro CLIA Plate (Dismountable) 96T: 8 wells ×12 strips strips -20℃, 6 months
Reference Standard 96T: 2 vials 48T: 1 vial -20℃, 6 months
Concentrated Biotinylated Detection Ab (100×) 96T: 1 vial, 120 μL 60 μL -20℃, 6 months
Concentrated HRP Conjugate (100×) 96T: 1 vial, 120 μL 60 μL -20℃ (Protect from light), 6 months
Reference Standard & Sample Diluent 1 vial, 20 mL 2-8°C, 6 months
Biotinylated Detection Ab Diluent 1 vial, 14 mL 2-8°C, 6 months
HRP Conjugate Diluent 1 vial, 14 mL 2-8°C, 6 months
Concentrated Wash Buffer (25×) 1 vial, 30 mL 2-8°C, 6 months
Substrate Reagent A 1 vial, 5 mL 2-8℃ (Protect from light)
Substrate Reagent B 1 vial, 5 mL 2-8℃ (Protect from light)
Plate Sealer 5 pieces
Manual 1 copy
Certificate of Analysis 1 copy

Sample Type Range (%) Average Recovery (%)
Serum(n=8) 94-109 101
EDTA plasma(n=8) 95-109 100
Cell culture media(n=8) 88-102 95

Intra-assay Precision Intra-assay Precision Intra-assay Precision Inter-assay Precision Inter-assay Precision Inter-assay Precision
Sample 1.00 2.00 3.00 1.00 2.00 3.00
n 20.00 20.00 20.00 20.00 20.00 20.00
Mean (pg/mL) 106.77 206.02 743.87 115.47 190.30 718.62
Standard deviation 10.84 15.39 51.48 10.38 17.20 81.13
CV (%) 10.15 7.47 6.92 8.99 9.04 11.29

Citations

Product Review

Well-cited Academic Publications
KO-Validated 700+ Lines
Technical Support Expert Scientific Team
1-Year Guarantee Antibodies & Proteins