Human IL10 (Interleukin-10) CLIA Kit for Quantitative Detection

SPECIFICATIONS
STJC0000093
🚚 Free UK Delivery on orders over £150
Processing The item has been added
Enquire For Bulk Order
✓ 1-Year Performance Guarantee | SDS & Datasheet Included | Expert Tech Support | Covered by St John's Laboratory Guarantee

General Information

Short DescriptionThis human IL10 kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes in serum, plasma and other biological fluids.
ApplicationsCLIA
ReactivityHuman
Sensitivity4.69pg/mL
Detection Limit7.81~500pg/mL
NoteFOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS.

Product Properties

Storage InstructionIf unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual.
Assay Time3.5h
DetectionChemiluminescence

Target Information

Gene SymbolIL10
Gene ID3586
Uniprot IDIL10_HUMAN
SpecificityThis kit recognizes Human IL-10 in samples. No significant cross-reactivity or interference between Human IL-10 and analogues was observed.
Sample TypeSerum, plasma and other biological fluids

Additional Info

Tissue Specificity Produced by a variety of cell lines, including T-cells, macrophages, mast cells and other cell types.
Function Major immune regulatory cytokine that acts on many cells of the immune system where it has profound anti-inflammatory functions, limiting excessive tissue disruption caused by inflammation. Mechanistically, IL10 binds to its heterotetrameric receptor comprising IL10RA and IL10RB leading to JAK1 and STAT2-mediated phosphorylation of STAT3. In turn, STAT3 translocates to the nucleus where it drives expression of anti-inflammatory mediators. Targets antigen-presenting cells (APCs) such as macrophages and monocytes and inhibits their release of pro-inflammatory cytokines including granulocyte-macrophage colony-stimulating factor /GM-CSF, granulocyte colony-stimulating factor/G-CSF, IL-1 alpha, IL-1 beta, IL-6, IL-8 and TNF-alpha. Also interferes with antigen presentation by reducing the expression of MHC-class II and co-stimulatory molecules, thereby inhibiting their ability to induce T cell activation. In addition, controls the inflammatory response of macrophages by reprogramming essential metabolic pathways including mTOR signaling.
Protein Name Interleukin-10
Il-10
Cytokine Synthesis Inhibitory Factor
Csif
Database Links Reactome: R-HSA-6783783
Reactome: R-HSA-6785807
Reactome: R-HSA-8950505
Reactome: R-HSA-9662834
Reactome: R-HSA-9664323
Reactome: R-HSA-9725370
Cellular Localisation Secreted
Alternative CLIA Names Interleukin-10 CLIA kit
Il-10 CLIA kit
Cytokine Synthesis Inhibitory Factor CLIA kit
Csif CLIA kit
IL10 CLIA kit
Specificity This kit recognizes Human IL-10 in samples. No significant cross-reactivity or interference between Human IL-10 and analogues was observed.
Reproducibility Both intra-CV and inter-CV are

Information sourced from Uniprot.org


Item Specifications Storage
Micro CLIA Plate (Dismountable) 96T: 8 wells ×12 strips strips -20℃, 6 months
Reference Standard 96T: 2 vials 48T: 1 vial -20℃, 6 months
Concentrated Biotinylated Detection Ab (100×) 96T: 1 vial, 120 μL 60 μL -20℃, 6 months
Concentrated HRP Conjugate (100×) 96T: 1 vial, 120 μL 60 μL -20℃ (Protect from light), 6 months
Reference Standard & Sample Diluent 1 vial, 20 mL 2-8°C, 6 months
Biotinylated Detection Ab Diluent 1 vial, 14 mL 2-8°C, 6 months
HRP Conjugate Diluent 1 vial, 14 mL 2-8°C, 6 months
Concentrated Wash Buffer (25×) 1 vial, 30 mL 2-8°C, 6 months
Substrate Reagent A 1 vial, 5 mL 2-8℃ (Protect from light)
Substrate Reagent B 1 vial, 5 mL 2-8℃ (Protect from light)
Plate Sealer 5 pieces
Manual 1 copy
Certificate of Analysis 1 copy

Sample Type Range (%) Average Recovery (%)
Serum(n=8) 92-106 98
EDTA plasma(n=8) 90-101 95
Cell culture media(n=8) 86-97 91

Intra-assay Precision Intra-assay Precision Intra-assay Precision Inter-assay Precision Inter-assay Precision Inter-assay Precision
Sample 1.00 2.00 3.00 1.00 2.00 3.00
n 20.00 20.00 20.00 20.00 20.00 20.00
Mean (pg/mL) 24.45 46.94 167.57 22.04 49.29 171.49
Standard deviation 3.13 5.51 16.12 2.52 5.52 13.31
CV (%) 12.80 11.74 9.62 11.43 11.20 7.76

Citations