Rat Mmp3 (Stromelysin-1) Sandwich ELISA Kit (STJE0005903)
SPECIFICATIONS
General Information
| Short Description | This Mmp3 Sandwich ELISA is an in-vitro enzyme-linked immunosorbent assay for the measurement of samples in rat serum, plasma and other biological fluids. |
| Applications | ELISA |
| Reactivity | Rat |
| Sensitivity | 0.09ng/mL |
| Detection Limit | 0.16~10ng/mL |
| Note | FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS. |
Product Properties
| Storage Instruction | If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual. |
| Assay Time | 3.5h |
| Detection | Colormetric |
Target Information
| Gene Symbol | Mmp3Uniprot ID=MMP3_RAT" |
| Specificity | This kit recognizes Rat MMP-3 in samples. No significant cross-reactivity or interference between Rat MMP-3 and analogues was observed. |
| Sample Type | Serum, plasma and other biological fluids |
Additional Info
| Protein Name | Stromelysin-1Sl-1Matrix Metalloproteinase-3Mmp-3Ptr1 ProteinTransin-1 |
| Cellular Localisation | SecretedExtracellular SpaceExtracellular Matrix |
| Alternative ELISA Names | Stromelysin-1 ELISA kitSl-1 ELISA kitMatrix Metalloproteinase-3 ELISA kitMmp-3 ELISA kitPtr1 Protein ELISA kitTransin-1 ELISA kitMmp3 ELISA kit |
| Specificity | This kit recognizes Rat MMP-3 in samples. No significant cross-reactivity or interference between Rat MMP-3 and analogues was observed. |
| Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
| Item | Specifications | Storage |
|---|---|---|
| Micro ELISA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
| Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
| Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
| Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
| Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
| Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
| HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
| Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
| Substrate Reagent | 1 vial, 10 mL | 2-8℃ (Protect from light) |
| Stop Solution | 1 vial, 10 mL | 2-8°C |
| Plate Sealer | 5 pieces | |
| Manual | 1 copy | |
| Certificate of Analysis | 1 copy |
| Sample Type | Range (%) | Average Recovery (%) |
|---|---|---|
| Serum(n=8) | 94-110 | 102 |
| EDTA plasma (n=8) | 85-96 | 90 |
| Cell culture media (n=8) | 85-98 | 92 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision | |
|---|---|---|---|---|---|---|
| Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
| n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
| Mean (ng/mL) | 0.50 | 1.10 | 4.30 | 0.50 | 1.10 | 4.40 |
| Standard deviation | 0.03 | 0.06 | 0.19 | 0.03 | 0.06 | 0.15 |
| CV (%) | 6.00 | 5.45 | 4.42 | 6.00 | 5.45 | 3.41 |