Mouse IgE (Immunoglobulin E) Sandwich ELISA Kit (STJE0005350)
SPECIFICATIONS
General Information
| Short Description | This IgE Sandwich ELISA is an in-vitro enzyme-linked immunosorbent assay for the measurement of samples in mouse serum, plasma and other biological fluids. |
| Applications | ELISA |
| Reactivity | Mouse |
| Sensitivity | 75ng/mL |
| Detection Limit | 125~8000ng/mL |
| Note | FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS. |
Product Properties
| Storage Instruction | If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual. |
| Assay Time | 3.5h |
| Detection | Colormetric |
Target Information
| Specificity | This kit recognizes Mouse IgE in samples. No significant cross-reactivity or interference between Mouse IgE and analogues was observed. |
| Sample Type | Serum, plasma and other biological fluids |
Additional Info
| Specificity | This kit recognizes Mouse IgE in samples. No significant cross-reactivity or interference between Mouse IgE and analogues was observed. |
| Reproducibility | Both intra-CV and inter-CV are < 10%. |
Information sourced from Uniprot.org
| Item | Specifications | Storage |
|---|---|---|
| Micro ELISA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
| Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
| Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
| Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
| Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
| Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
| HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
| Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
| Substrate Reagent | 1 vial, 10 mL | 2-8℃ (Protect from light) |
| Stop Solution | 1 vial, 10 mL | 2-8°C |
| Plate Sealer | 5 pieces | |
| Manual | 1 copy | |
| Certificate of Analysis | 1 copy |
| Sample Type | Range (%) | Average Recovery (%) |
|---|---|---|
| Serum(n=8) | 101-111 | 104 |
| EDTA plasma (n=8) | 86-97 | 93 |
| Cell culture media (n=8) | 94-108 | 102 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision | |
|---|---|---|---|---|---|---|
| Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
| n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
| Mean (ng/mL) | 382.47 | 858.80 | 3319.11 | 351.27 | 946.20 | 3588.32 |
| Standard deviation | 21.34 | 40.54 | 153.34 | 19.81 | 46.65 | 191.98 |
| CV (%) | 5.58 | 4.72 | 4.62 | 5.64 | 4.93 | 5.35 |