Applications: |
CLIA |
Reactivity: |
Human |
Note: |
FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS. |
Sensitivity: |
7.5pg/mL |
Detection Limit: |
12.5~800pg/mL |
Short Description: |
This human NEFL kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes. |
Storage Instruction: |
If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual. |
Assay Time: |
3.5h |
Detection: |
Chemiluminescence |
Gene Symbol: |
NEFL |
Gene ID: |
4747 |
Uniprot ID: |
NFL_HUMAN |
Specificity: |
This kit recognizes Human NEFL in samples. No significant cross-reactivity or interference between Human NEFL and analogues was observed. |
Sample Type: |
Serum, plasma and other biological fluids |
Post Translational Modifications | O-glycosylated. Phosphorylated in the head and rod regions by the PKC kinase PKN1, leading to the inhibition of polymerization. Ubiquitinated in the presence of TRIM2 and UBE2D1. |
Function | Neurofilaments usually contain three intermediate filament proteins: NEFL, NEFM, and NEFH which are involved in the maintenance of neuronal caliber. May additionally cooperate with the neuronal intermediate filament proteins PRPH and INA to form neuronal filamentous networks. |
Protein Name | Neurofilament Light PolypeptideNf-L68 Kda Neurofilament ProteinNeurofilament Triplet L Protein |
Database Links | Reactome: R-HSA-438066Reactome: R-HSA-442982Reactome: R-HSA-5673001Reactome: R-HSA-9609736Reactome: R-HSA-9617324Reactome: R-HSA-9620244 |
Cellular Localisation | Cell ProjectionAxonCytoplasmCytoskeleton |
Alternative CLIA Names | Neurofilament Light Polypeptide CLIA kitNf-L CLIA kit68 Kda Neurofilament Protein CLIA kitNeurofilament Triplet L Protein CLIA kitNEFL CLIA kitNF68 CLIA kitNFL CLIA kit |
Specificity | This kit recognizes Human NEFL in samples. No significant cross-reactivity or interference between Human NEFL and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
Item | Specifications | Storage |
Micro CLIA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
Substrate Reagent A | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Substrate Reagent B | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Plate Sealer | 5 pieces | |
Manual | 1 copy | |
Certificate of Analysis | 1 copy | |
Sample Type | Range (%) | Average Recovery (%) |
Serum(n=8) | 95-109 | 100 |
EDTA plasma(n=8) | 96-111 | 104 |
Cell culture media(n=8) | 97-108 | 103 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision |
Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
Mean (pg/mL) | 42.81 | 110.42 | 359.68 | 45.57 | 109.06 | 388.00 |
Standard deviation | 4.76 | 8.26 | 23.49 | 5.50 | 13.02 | 28.75 |
CV (%) | 11.12 | 7.48 | 6.53 | 12.07 | 11.94 | 7.41 |
12 months for antibodies. 6 months for ELISA Kits. Please see website T&Cs for further guidance