| Specificity | This kit recognizes Mouse IL-12 in samples. No significant cross-reactivity or interference between Mouse IL-12 and analogues was observed. | 
  | Reproducibility | Both intra-CV and inter-CV are | 
 
Information sourced from Uniprot.org
   | Item | Specifications | Storage | 
  | Micro CLIA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months | 
  | Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months | 
  | Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months | 
  | Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months | 
  | Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months | 
  | Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months | 
  | HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months | 
  | Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months | 
  | Substrate Reagent A | 1 vial, 5 mL | 2-8℃ (Protect from light) | 
  | Substrate Reagent B | 1 vial, 5 mL | 2-8℃ (Protect from light) | 
  | Plate Sealer | 5 pieces |  | 
  | Manual | 1 copy |  | 
  | Certificate of Analysis | 1 copy |  | 
 
   | Sample Type | Range (%) | Average Recovery (%) | 
  | Serum(n=8) | 98-115 | 106 | 
  | EDTA plasma(n=8) | 83-97 | 90 | 
  | Cell culture media(n=8) | 94-110 | 101 | 
 
   |  | Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision | 
  | Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 | 
  | n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 
  | Mean (pg/mL) | 25.83 | 67.54 | 196.87 | 24.12 | 70.00 | 178.39 | 
  | Standard deviation | 2.29 | 6.25 | 15.16 | 2.32 | 7.94 | 16.80 | 
  | CV (%) | 8.87 | 9.25 | 7.70 | 9.62 | 11.34 | 9.42 |