Specificity | This kit recognizes IgG3 in samples. No significant cross-reactivity or interference between IgG3 and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
Item | Specifications | Storage |
Micro ELISA Plate (Dismountable) | 96T: 8 wells ×12 strips strips 24T: 8 wells ×3 strips 96T*5: 5 plates, 96T | 2-8℃, 1 months |
Reference Standard | 96T: 2 vials 48T/24T: 1 vial 96T*5: 10 vials | 2-8℃, use the reconstituted standard within 24h |
Concentrated HRP Linked Detection Ab (100×) | 96T: 1 vial, 60 μL 48T/24T: 1 vial, 30 μL 96T*5: 5 vials, 60 μL | 2-8℃ (Protect from light) |
Reference Standard & Sample Diluent | 96T/48T/24T: 1 vial, 20 mL96T*5: 5 vials, 20 mL | 2-8℃ |
HRP Linked Ab Diluent | 96T/48T/24T: 1 vial, 14 mL96T*5: 5 vials, 14 mL | |
Concentrated Wash Buffer (25×) | 96T/48T/24T: 1 vial, 30 mL96T*5: 5 vials, 30 mL | |
Substrate Reagent | 96T/48T/24T: 1 vial, 10 mL96T*5: 5 vials, 10 mL | 2-8℃ (Protect from light) |
Stop Solution | 96T/48T/24T: 1 vial, 10 mL96T*5: 5 vials, 10 mL | 2-8℃ |
Plate Sealer | 96T/48T/24T: 5 pieces96T*5: 25 pieces | |
Product Description | 1 copy | |
Certificate of Analysis | 1 copy | |
Sample Type | Range (%) | Average Recovery (%) |
Serum(n=8) | 86-101 | 92 |
EDTA plasma(n=8) | 91-101 | 96 |
Cell culture media(n=8) | 93-108 | 101 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision |
Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
Mean (ng/mL) | 1.02 | 2.84 | 7.14 | 1.09 | 3.02 | 6.65 |
Standard deviation | 0.07 | 0.16 | 0.29 | 0.06 | 0.18 | 0.27 |
CV (%) | 6.77 | 5.46 | 4.08 | 5.83 | 5.90 | 4.02 |