Specificity | This kit recognizes Human IL-12 in samples. No significant cross-reactivity or interference between Human IL-12 and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
Item | Specifications | Storage |
Micro CLIA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
Substrate Reagent A | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Substrate Reagent B | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Plate Sealer | 5 pieces | |
Manual | 1 copy | |
Certificate of Analysis | 1 copy | |
Sample Type | Range (%) | Average Recovery (%) |
Serum(n=8) | 90-104 | 98 |
EDTA plasma(n=8) | 92-105 | 98 |
Cell culture media(n=8) | 100-114 | 106 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision |
Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
Mean (pg/mL) | 41.75 | 101.46 | 316.46 | 43.49 | 101.51 | 318.17 |
Standard deviation | 4.46 | 9.65 | 29.49 | 4.00 | 9.16 | 34.24 |
CV (%) | 10.68 | 9.51 | 9.32 | 9.20 | 9.02 | 10.76 |