Applications: | CLIA |
Reactivity: | Human |
Note: | FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS. |
Sensitivity: | 9.38pg/mL |
Detection Limit: | 15.63~1000pg/mL |
Short Description : | This human CDH3 kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes in serum, plasma and other biological fluids. |
Storage Instruction: | If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual. |
Assay Time: | 3.5h |
Detection: | Chemiluminescence |
Gene Symbol: | CDH3 |
Gene ID: | 1001 |
Uniprot ID: | CADH3_HUMAN |
Specificity: | This kit recognizes Human P-Cadherin in samples. No significant cross-reactivity or interference between Human P-Cadherin and analogues was observed. |
Sample Type: | Serum, plasma and other biological fluids |
Tissue Specificity | Expressed in some normal epithelial tissues and in some carcinoma cell lines. |
Function | Cadherins are calcium-dependent cell adhesion proteins. They preferentially interact with themselves in a homophilic manner in connecting cells.cadherins may thus contribute to the sorting of heterogeneous cell types. |
Protein Name | Cadherin-3 Placental Cadherin P-Cadherin |
Database Links | Reactome: R-HSA-418990 |
Cellular Localisation | Cell Membrane Single-Pass Type I Membrane Protein |
Alternative CLIA Names | Cadherin-3 CLIA kit Placental Cadherin CLIA kit P-Cadherin CLIA kit CDH3 CLIA kit CDHP CLIA kit |
Specificity | This kit recognizes Human P-Cadherin in samples. No significant cross-reactivity or interference between Human P-Cadherin and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
Item | Specifications | Storage |
---|---|---|
Micro CLIA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
Substrate Reagent A | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Substrate Reagent B | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Plate Sealer | 5 pieces | |
Manual | 1 copy | |
Certificate of Analysis | 1 copy |
Sample Type | Range (%) | Average Recovery (%) |
---|---|---|
Serum(n=8) | 99-116 | 107 |
EDTA plasma(n=8) | 91-106 | 98 |
Cell culture media(n=8) | 98-112 | 105 |
Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision | |
---|---|---|---|---|---|---|
Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
Mean (pg/mL) | 52.30 | 132.83 | 408.68 | 56.36 | 141.50 | 442.94 |
Standard deviation | 4.69 | 10.80 | 25.09 | 5.28 | 9.92 | 27.64 |
CV (%) | 8.97 | 8.13 | 6.14 | 9.37 | 7.01 | 6.24 |