Applications: |
CLIA |
Reactivity: |
Human |
Note: |
FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS. |
Sensitivity: |
37.5pg/mL |
Detection Limit: |
62.5~4000pg/mL |
Short Description: |
This human CA12 kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes. |
Storage Instruction: |
If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual. |
Assay Time: |
3.5h |
Detection: |
Chemiluminescence |
Gene Symbol: |
CA12 |
Gene ID: |
771 |
Uniprot ID: |
CAH12_HUMAN |
Specificity: |
This kit recognizes Human CA12 in samples. No significant cross-reactivity or interference between Human CA12 and analogues was observed. |
Sample Type: |
Serum, plasma and other biological fluids |
Tissue Specificity | Highly expressed in colon, kidney, prostate, intestine and activated lymphocytes. Expressed at much higher levels in the renal cell cancers than in surrounding normal kidney tissue. Moderately expressed in pancreas, ovary and testis. Expressed in sweat glands and bronchiolar epithelium. |
Function | Reversible hydration of carbon dioxide. |
Protein Name | Carbonic Anhydrase 12Carbonate Dehydratase XiiCarbonic Anhydrase XiiCa-XiiTumor Antigen Hom-Rcc-3.1.3 |
Database Links | Reactome: R-HSA-1475029 |
Cellular Localisation | MembraneSingle-Pass Type I Membrane ProteinCell Membrane |
Alternative CLIA Names | Carbonic Anhydrase 12 CLIA kitCarbonate Dehydratase Xii CLIA kitCarbonic Anhydrase Xii CLIA kitCa-Xii CLIA kitTumor Antigen Hom-Rcc-3.1.3 CLIA kitCA12 CLIA kit |
Specificity | This kit recognizes Human CA12 in samples. No significant cross-reactivity or interference between Human CA12 and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
Item | Specifications | Storage |
Micro CLIA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
Substrate Reagent A | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Substrate Reagent B | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Plate Sealer | 5 pieces | |
Manual | 1 copy | |
Certificate of Analysis | 1 copy | |
Sample Type | Range (%) | Average Recovery (%) |
Serum(n=8) | 88-102 | 93 |
EDTA plasma(n=8) | 85-99 | 92 |
Cell culture media(n=8) | 89-102 | 97 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision |
Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
Mean (pg/mL) | 188.46 | 452.17 | 1616.46 | 183.05 | 457.51 | 1483.26 |
Standard deviation | 23.75 | 50.91 | 148.39 | 17.52 | 48.36 | 101.01 |
CV (%) | 12.60 | 11.26 | 9.18 | 9.57 | 10.57 | 6.81 |
12 months for antibodies. 6 months for ELISA Kits. Please see website T&Cs for further guidance