• Human CA3 (Carbonic anhydrase 3) CLIA Kit (STJC0000219)

Human CA3 (Carbonic anhydrase 3) CLIA Kit (STJC0000219)

SKU:
STJC0000219

Shipping:
Free Shipping
Current Stock:
Applications: CLIA
Reactivity: Human
Note: FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS.
Sensitivity: 0.38ng/mL
Detection Limit: 0.63~40ng/mL
Short Description: This human CA3 kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes.
Storage Instruction: If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual.
Assay Time: 3.5h
Detection: Chemiluminescence
Gene Symbol: CA3
Gene ID: 761
Uniprot ID: CAH3_HUMAN
Specificity: This kit recognizes Human CA3 in samples. No significant cross-reactivity or interference between Human CA3 and analogues was observed.
Sample Type: Serum, plasma and other biological fluids
Tissue Specificity Muscle specific.
Post Translational Modifications S-thiolated both by thiol-disulfide exchange with glutathione disulfide and by oxyradical-initiated S-thiolation with reduced glutathione. S-glutathionylated in hepatocytes under oxidative stress.
Function Reversible hydration of carbon dioxide.
Protein Name Carbonic Anhydrase 3
Carbonate Dehydratase Iii
Carbonic Anhydrase Iii
Ca-Iii
Database Links Reactome: R-HSA-1475029
Cellular Localisation Cytoplasm
Alternative CLIA Names Carbonic Anhydrase 3 CLIA kit
Carbonate Dehydratase Iii CLIA kit
Carbonic Anhydrase Iii CLIA kit
Ca-Iii CLIA kit
CA3 CLIA kit
Specificity This kit recognizes Human CA3 in samples. No significant cross-reactivity or interference between Human CA3 and analogues was observed.
Reproducibility Both intra-CV and inter-CV are

Information sourced from Uniprot.org


Item Specifications Storage
Micro CLIA Plate (Dismountable) 96T: 8 wells ×12 strips strips -20℃, 6 months
Reference Standard 96T: 2 vials 48T: 1 vial -20℃, 6 months
Concentrated Biotinylated Detection Ab (100×) 96T: 1 vial, 120 μL 60 μL -20℃, 6 months
Concentrated HRP Conjugate (100×) 96T: 1 vial, 120 μL 60 μL -20℃ (Protect from light), 6 months
Reference Standard & Sample Diluent 1 vial, 20 mL 2-8°C, 6 months
Biotinylated Detection Ab Diluent 1 vial, 14 mL 2-8°C, 6 months
HRP Conjugate Diluent 1 vial, 14 mL 2-8°C, 6 months
Concentrated Wash Buffer (25×) 1 vial, 30 mL 2-8°C, 6 months
Substrate Reagent A 1 vial, 5 mL 2-8℃ (Protect from light)
Substrate Reagent B 1 vial, 5 mL 2-8℃ (Protect from light)
Plate Sealer 5 pieces
Manual 1 copy
Certificate of Analysis 1 copy

Sample Type Range (%) Average Recovery (%)
Serum(n=8) 99-114 105
EDTA plasma(n=8) 99-115 106
Cell culture media(n=8) 101-116 107

Intra-assay Precision Intra-assay Precision Intra-assay Precision Inter-assay Precision Inter-assay Precision Inter-assay Precision
Sample 1.00 2.00 3.00 1.00 2.00 3.00
n 20.00 20.00 20.00 20.00 20.00 20.00
Mean (ng/mL) 1.95 3.50 17.49 1.85 3.63 18.82
Standard deviation 0.17 0.26 1.81 0.16 0.29 1.50
CV (%) 8.72 7.43 10.35 8.65 7.99 7.97
12 months for antibodies. 6 months for ELISA Kits. Please see website T&Cs for further guidance