Applications: |
CLIA |
Reactivity: |
Rat |
Note: |
FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS. |
Sensitivity: |
18.75pg/mL |
Detection Limit: |
31.25~2000pg/mL |
Short Description: |
This rat HO1 kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes. |
Storage Instruction: |
If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual. |
Assay Time: |
3.5h |
Detection: |
Chemiluminescence |
Gene Symbol: |
Hmox1 |
Gene ID: |
24451 |
Uniprot ID: |
HMOX1_RAT |
Specificity: |
This kit recognizes Rat HO1 in samples. No significant cross-reactivity or interference between Rat HO1 and analogues was observed. |
Sample Type: |
Serum, plasma and other biological fluids |
Post Translational Modifications | A soluble form arises by proteolytic removal of the membrane anchor. |
Function | Heme oxygenase 1: Catalyzes the oxidative cleavage of heme at the alpha-methene bridge carbon, released as carbon monoxide (CO), to generate biliverdin IXalpha, while releasing the central heme iron chelate as ferrous iron. Affords protection against programmed cell death and this cytoprotective effect relies on its ability to catabolize free heme and prevent it from sensitizing cells to undergo apoptosis. Heme oxygenase 1 soluble form: Catalyzes the oxidative cleavage of heme at the alpha-methene bridge carbon, released as carbon monoxide (CO), to generate biliverdin IXalpha, while releasing the central heme iron chelate as ferrous iron. |
Protein Name | Heme Oxygenase 1Ho-1Hsp32 Cleaved Into - Heme Oxygenase 1 Soluble Form |
Database Links | Reactome: R-RNO-189483Reactome: -RNO-917937Reactome: -RNO-9609523Reactome: -RNO-9707564Reactome: -RNO-9707587 |
Cellular Localisation | Endoplasmic Reticulum MembraneSingle-Pass Type Iv Membrane ProteinCytoplasmic Side |
Alternative CLIA Names | Heme Oxygenase 1 CLIA kitHo-1 CLIA kitHsp32 Cleaved Into - Heme Oxygenase 1 Soluble Form CLIA kitHmox1 CLIA kit |
Specificity | This kit recognizes Rat HO1 in samples. No significant cross-reactivity or interference between Rat HO1 and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
Item | Specifications | Storage |
Micro CLIA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
Substrate Reagent A | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Substrate Reagent B | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Plate Sealer | 5 pieces | |
Manual | 1 copy | |
Certificate of Analysis | 1 copy | |
Sample Type | Range (%) | Average Recovery (%) |
Serum(n=8) | 87-100 | 94 |
EDTA plasma(n=8) | 100-112 | 106 |
Cell culture media(n=8) | 84-97 | 91 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision |
Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
Mean (pg/mL) | 103.92 | 212.77 | 683.28 | 100.13 | 194.89 | 751.37 |
Standard deviation | 9.81 | 18.60 | 44.89 | 12.94 | 14.99 | 81.00 |
CV (%) | 9.44 | 8.74 | 6.57 | 12.92 | 7.69 | 10.78 |
12 months for antibodies. 6 months for ELISA Kits. Please see website T&Cs for further guidance