Applications: |
ELISA |
Reactivity: |
Human |
Note: |
FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS. |
Sensitivity: |
1.03 pg/mL |
Detection Limit: |
3.13-200 pg/mL |
Short Description: |
This NEFL Sandwich ELISA is an in-vitro enzyme-linked immunosorbent assay for the measurement of samples in human cell culture supernatant, serum and plasma (EDTA, citrate, heparin). |
Storage Instruction: |
If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual. |
Assay Time: |
3.5h |
Detection: |
Colormetric |
Gene Symbol: |
NEFL |
Gene ID: |
4747 |
Uniprot ID: |
NFL_HUMAN |
Specificity: |
This kit recognizes Human NEFL in samples. No significant cross-reactivity or interference between Human NEFL and analogues was observed. |
Sample Type: |
serum, plasma and other biological fluids |
Post Translational Modifications | O-glycosylated. Phosphorylated in the head and rod regions by the PKC kinase PKN1, leading to the inhibition of polymerization. Ubiquitinated in the presence of TRIM2 and UBE2D1. |
Function | Neurofilaments usually contain three intermediate filament proteins: NEFL, NEFM, and NEFH which are involved in the maintenance of neuronal caliber. May additionally cooperate with the neuronal intermediate filament proteins PRPH and INA to form neuronal filamentous networks. |
Protein Name | Neurofilament Light PolypeptideNf-L68 Kda Neurofilament ProteinNeurofilament Triplet L Protein |
Database Links | Reactome: R-HSA-438066Reactome: R-HSA-442982Reactome: R-HSA-5673001Reactome: R-HSA-9609736Reactome: R-HSA-9617324Reactome: R-HSA-9620244 |
Cellular Localisation | Cell ProjectionAxonCytoplasmCytoskeleton |
Alternative ELISA Names | Neurofilament Light Polypeptide ELISA kitNf-L ELISA kit68 Kda Neurofilament Protein ELISA kitNeurofilament Triplet L Protein ELISA kitNEFL ELISA kitNF68 ELISA kitNFL ELISA kit |
Specificity | This kit recognizes Human NEFL in samples. No significant cross-reactivity or interference between Human NEFL and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
Item | Specifications | Storage |
Micro ELISA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
Substrate Reagent | 1 vial, 10 mL | 2-8℃ (Protect from light) |
Stop Solution | 1 vial, 10 mL | 2-8°C |
Plate Sealer | 5 pieces | |
Manual | 1 copy | |
Certificate of Analysis | 1 copy | |
Sample Type | Range (%) | Average Recovery (%) |
Serum(n=8) | 84-99 | 91 |
EDTA plasma (n=8) | 87-99 | 92 |
Cell culture media (n=8) | 91-102 | 97 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision |
Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
Mean (pg/mL) | 9.31 | 21.34 | 71.13 | 9.72 | 20.50 | 74.21 |
Standard deviation | 0.58 | 1.02 | 3.31 | 0.57 | 1.17 | 2.72 |
CV (%) | 6.22 | 4.80 | 4.66 | 5.85 | 5.72 | 3.66 |
12 months for antibodies. 6 months for ELISA Kits. Please see website T&Cs for further guidance