Applications: |
ELISA |
Reactivity: |
Human |
Note: |
FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS. |
Sensitivity: |
0.09ng/mL |
Detection Limit: |
0.16~10ng/mL |
Short Description: |
This NPHS1 Sandwich ELISA is an in-vitro enzyme-linked immunosorbent assay for the measurement of samples in human cell culture supernatant, serum and plasma (EDTA, citrate, heparin). |
Storage Instruction: |
If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual. |
Assay Time: |
3.5h |
Detection: |
Colormetric |
Gene Symbol: |
NPHS1 |
Gene ID: |
4868 |
Uniprot ID: |
NPHN_HUMAN |
Specificity: |
This kit recognizes Human NPHN in samples. No significant cross-reactivity or interference between Human NPHN and analogues was observed. |
Sample Type: |
Serum, plasma and other biological fluids |
Post Translational Modifications | Phosphorylated at Tyr-1193 by FYN, leading to the recruitment and activation of phospholipase C-gamma-1/PLCG1. Tyrosine phosphorylation is reduced by high glucose levels. Dephosphorylated by tensin TNS2 which leads to reduced binding of NPHN1 to PIK3R1. |
Function | Seems to play a role in the development or function of the kidney glomerular filtration barrier. Regulates glomerular vascular permeability. May anchor the podocyte slit diaphragm to the actin cytoskeleton. Plays a role in skeletal muscle formation through regulation of myoblast fusion. |
Protein Name | NephrinRenal Glomerulus-Specific Cell Adhesion Receptor |
Database Links | Reactome: R-HSA-373753 |
Cellular Localisation | Cell MembraneSingle-Pass Type I Membrane ProteinPredominantly Located At Podocyte Slit Diaphragm Between Podocyte Foot ProcessesAlso Associated With Podocyte Apical Plasma Membrane |
Alternative ELISA Names | Nephrin ELISA kitRenal Glomerulus-Specific Cell Adhesion Receptor ELISA kitNPHS1 ELISA kitNPHN ELISA kit |
Specificity | This kit recognizes Human NPHN in samples. No significant cross-reactivity or interference between Human NPHN and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
Item | Specifications | Storage |
Micro ELISA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
Substrate Reagent | 1 vial, 10 mL | 2-8℃ (Protect from light) |
Stop Solution | 1 vial, 10 mL | 2-8°C |
Plate Sealer | 5 pieces | |
Manual | 1 copy | |
Certificate of Analysis | 1 copy | |
Sample Type | Range (%) | Average Recovery (%) |
Serum(n=8) | 96-108 | 101 |
EDTA plasma (n=8) | 90-106 | 96 |
Cell culture media (n=8) | 89-102 | 95 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision |
Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
Mean (ng/mL) | 0.48 | 0.90 | 3.58 | 0.52 | 0.83 | 3.85 |
Standard deviation | 0.03 | 0.05 | 0.19 | 0.03 | 0.03 | 0.18 |
CV (%) | 6.25 | 5.56 | 5.31 | 5.77 | 3.61 | 4.68 |
12 months for antibodies. 6 months for ELISA Kits. Please see website T&Cs for further guidance