Applications: |
CLIA |
Reactivity: |
Human |
Note: |
FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS. |
Sensitivity: |
0.38ng/mL |
Detection Limit: |
0.63~40ng/mL |
Short Description: |
This human GP6 kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes. |
Storage Instruction: |
If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual. |
Assay Time: |
3.5h |
Detection: |
Chemiluminescence |
Gene Symbol: |
GP6 |
Gene ID: |
51206 |
Uniprot ID: |
GPVI_HUMAN |
Specificity: |
This kit recognizes Human GP6 in samples. No significant cross-reactivity or interference between Human GP6 and analogues was observed. |
Sample Type: |
Serum, plasma and other biological fluids |
Tissue Specificity | Megakaryocytes and platelets. |
Post Translational Modifications | N-linked glycosylation at Asn-92 is not required for the cell surface expression, but contributes to maximal adhesion to type I collagen, collagen-related peptide (CRP), and, to a lesser extent, to the snake venom C-type lectin convulxin (CVX). |
Function | Collagen receptor involved in collagen-induced platelet adhesion and activation. Plays a key role in platelet procoagulant activity and subsequent thrombin and fibrin formation. This procoagulant function may contribute to arterial and venous thrombus formation. The signaling pathway involves the FcR gamma-chain, the Src kinases (likely FYN or LYN) and SYK, the adapter protein LAT and leads to the activation of PLCG2. |
Protein Name | Platelet Glycoprotein ViGpviGlycoprotein 6 |
Database Links | Reactome: R-HSA-114604Reactome: R-HSA-202733Reactome: R-HSA-75892 |
Cellular Localisation | Isoform 1: Cell MembraneSingle-Pass Membrane ProteinIsoform 2: Cell Membrane |
Alternative CLIA Names | Platelet Glycoprotein Vi CLIA kitGpvi CLIA kitGlycoprotein 6 CLIA kitGP6 CLIA kit |
Specificity | This kit recognizes Human GP6 in samples. No significant cross-reactivity or interference between Human GP6 and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
Item | Specifications | Storage |
Micro CLIA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
Substrate Reagent A | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Substrate Reagent B | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Plate Sealer | 5 pieces | |
Manual | 1 copy | |
Certificate of Analysis | 1 copy | |
Sample Type | Range (%) | Average Recovery (%) |
Serum(n=8) | 96-110 | 102 |
EDTA plasma(n=8) | 98-113 | 106 |
Cell culture media(n=8) | 94-106 | 100 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision |
Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
Mean (ng/mL) | 2.05 | 4.57 | 16.89 | 1.85 | 4.19 | 15.73 |
Standard deviation | 0.22 | 0.40 | 1.78 | 0.19 | 0.39 | 1.37 |
CV (%) | 10.73 | 8.75 | 10.54 | 10.27 | 9.31 | 8.71 |
12 months for antibodies. 6 months for ELISA Kits. Please see website T&Cs for further guidance