Applications: |
CLIA |
Reactivity: |
Rat |
Note: |
FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS. |
Sensitivity: |
4.69pg/mL |
Detection Limit: |
7.81~500pg/mL |
Short Description: |
This rat REN kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes. |
Storage Instruction: |
If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual. |
Assay Time: |
3.5h |
Detection: |
Chemiluminescence |
Gene Symbol: |
Ren1 |
Gene ID: |
24715 |
Uniprot ID: |
RENI_RAT |
Specificity: |
This kit recognizes Rat REN in samples. No significant cross-reactivity or interference between Rat REN and analogues was observed. |
Sample Type: |
Serum, plasma and other biological fluids |
Function | Renin is a highly specific endopeptidase, whose only known function is to generate angiotensin I from angiotensinogen in the plasma, initiating a cascade of reactions that produce an elevation of blood pressure and increased sodium retention by the kidney. |
Protein Name | ReninAngiotensinogenase |
Database Links | Reactome: R-RNO-2022377 |
Cellular Localisation | SecretedMembraneAssociated To Membranes Via Binding To Atp6ap2 |
Alternative CLIA Names | Renin CLIA kitAngiotensinogenase CLIA kitRen1 CLIA kitRen CLIA kit |
Specificity | This kit recognizes Rat REN in samples. No significant cross-reactivity or interference between Rat REN and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
Item | Specifications | Storage |
Micro CLIA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
Substrate Reagent A | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Substrate Reagent B | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Plate Sealer | 5 pieces | |
Manual | 1 copy | |
Certificate of Analysis | 1 copy | |
Sample Type | Range (%) | Average Recovery (%) |
Serum(n=8) | 98-110 | 103 |
EDTA plasma(n=8) | 96-110 | 103 |
Cell culture media(n=8) | 86-98 | 93 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision |
Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
Mean (pg/mL) | 24.24 | 42.13 | 218.10 | 22.99 | 45.88 | 224.12 |
Standard deviation | 1.99 | 3.30 | 23.93 | 2.40 | 4.81 | 18.15 |
CV (%) | 8.21 | 7.83 | 10.97 | 10.44 | 10.48 | 8.10 |