Applications: |
CLIA |
Reactivity: |
Human |
Note: |
FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS. |
Sensitivity: |
37.5pg/mL |
Detection Limit: |
62.5~4000pg/mL |
Short Description: |
This human TACR1 kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes. |
Storage Instruction: |
If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual. |
Assay Time: |
3.5h |
Detection: |
Chemiluminescence |
Gene Symbol: |
TACR1 |
Gene ID: |
6869 |
Uniprot ID: |
NK1R_HUMAN |
Specificity: |
This kit recognizes Human TACR1 in samples. No significant cross-reactivity or interference between Human TACR1 and analogues was observed. |
Sample Type: |
Serum, plasma and other biological fluids |
Function | This is a receptor for the tachykinin neuropeptide substance P. It is probably associated with G proteins that activate a phosphatidylinositol-calcium second messenger system. The rank order of affinity of this receptor to tachykinins is: substance P > substance K > neuromedin-K. |
Protein Name | Substance-P ReceptorSprNk-1 ReceptorNk-1rTachykinin Receptor 1 |
Database Links | Reactome: R-HSA-380095Reactome: R-HSA-416476Reactome: R-HSA-8856825Reactome: R-HSA-8856828 |
Cellular Localisation | Cell MembraneMulti-Pass Membrane Protein |
Alternative CLIA Names | Substance-P Receptor CLIA kitSpr CLIA kitNk-1 Receptor CLIA kitNk-1r CLIA kitTachykinin Receptor 1 CLIA kitTACR1 CLIA kitNK1R CLIA kitTAC1R CLIA kit |
Specificity | This kit recognizes Human TACR1 in samples. No significant cross-reactivity or interference between Human TACR1 and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
Item | Specifications | Storage |
Micro CLIA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
Substrate Reagent A | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Substrate Reagent B | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Plate Sealer | 5 pieces | |
Manual | 1 copy | |
Certificate of Analysis | 1 copy | |
Sample Type | Range (%) | Average Recovery (%) |
Serum(n=8) | 97-114 | 105 |
EDTA plasma(n=8) | 100-113 | 106 |
Cell culture media(n=8) | 86-96 | 91 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision |
Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
Mean (pg/mL) | 204.46 | 577.42 | 1395.25 | 222.35 | 609.64 | 1509.49 |
Standard deviation | 24.45 | 63.57 | 93.48 | 22.37 | 46.45 | 129.97 |
CV (%) | 11.96 | 11.01 | 6.70 | 10.06 | 7.62 | 8.61 |
12 months for antibodies. 6 months for ELISA Kits. Please see website T&Cs for further guidance