Applications: |
CLIA |
Reactivity: |
Human |
Note: |
FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS. |
Sensitivity: |
37.5pg/mL |
Detection Limit: |
62.5~4000pg/mL |
Short Description: |
This human SPARCL1 kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes. |
Storage Instruction: |
If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual. |
Assay Time: |
3.5h |
Detection: |
Chemiluminescence |
Gene Symbol: |
SPARCL1 |
Gene ID: |
8404 |
Uniprot ID: |
SPRL1_HUMAN |
Specificity: |
This kit recognizes Human SPARCL1 in samples. No significant cross-reactivity or interference between Human SPARCL1 and analogues was observed. |
Sample Type: |
Serum, plasma and other biological fluids |
Tissue Specificity | Highly expressed in lymph node, brain, heart, lung, skeletal muscle, ovary, small intestine, and colon, with lower levels in placenta, pancreas, testis, spleen, and thymus, and no expression in kidney, liver, and peripheral blood leukocytes. |
Post Translational Modifications | N- and O-glycosylated. O-glycosylated with a core 1 or possibly core 8 glycan. |
Protein Name | Sparc-Like Protein 1High Endothelial Venule ProteinHevinMast 9 |
Database Links | Reactome: R-HSA-381426Reactome: R-HSA-8957275 |
Cellular Localisation | SecretedExtracellular SpaceExtracellular Matrix |
Alternative CLIA Names | Sparc-Like Protein 1 CLIA kitHigh Endothelial Venule Protein CLIA kitHevin CLIA kitMast 9 CLIA kitSPARCL1 CLIA kit |
Specificity | This kit recognizes Human SPARCL1 in samples. No significant cross-reactivity or interference between Human SPARCL1 and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
Item | Specifications | Storage |
Micro CLIA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
Substrate Reagent A | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Substrate Reagent B | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Plate Sealer | 5 pieces | |
Manual | 1 copy | |
Certificate of Analysis | 1 copy | |
Sample Type | Range (%) | Average Recovery (%) |
Serum(n=8) | 98-115 | 106 |
EDTA plasma(n=8) | 88-102 | 95 |
Cell culture media(n=8) | 92-106 | 99 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision |
Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
Mean (pg/mL) | 197.73 | 553.13 | 1740.74 | 182.80 | 540.30 | 1831.46 |
Standard deviation | 21.08 | 60.01 | 105.31 | 18.72 | 59.32 | 149.81 |
CV (%) | 10.66 | 10.85 | 6.05 | 10.24 | 10.98 | 8.18 |
12 months for antibodies. 6 months for ELISA Kits. Please see website T&Cs for further guidance