Sensitivity: 37.5pg/mL
Detection Limit: 62.5~4000pg/mL
Short Description: This human SPARCL1 kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes.
Storage Instruction: If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual.
Assay Time: 3.5h
Detection: Chemiluminescence
Gene Symbol: SPARCL1
Gene ID: 8404
Uniprot ID: SPRL1_HUMAN
Specificity: This kit recognizes Human SPARCL1 in samples. No significant cross-reactivity or interference between Human SPARCL1 and analogues was observed.
Sample Type: Serum, plasma and other biological fluids
Tissue Specificity Highly expressed in lymph node, brain, heart, lung, skeletal muscle, ovary, small intestine, and colon, with lower levels in placenta, pancreas, testis, spleen, and thymus, and no expression in kidney, liver, and peripheral blood leukocytes.
Post Translational Modifications N- and O-glycosylated. O-glycosylated with a core 1 or possibly core 8 glycan.
Protein Name Sparc-Like Protein 1
High Endothelial Venule Protein
Hevin
Mast 9
Database Links Reactome: R-HSA-381426
Reactome: R-HSA-8957275
Cellular Localisation Secreted
Extracellular Space
Extracellular Matrix
Alternative CLIA Names Sparc-Like Protein 1 CLIA kit
High Endothelial Venule Protein CLIA kit
Hevin CLIA kit
Mast 9 CLIA kit
SPARCL1 CLIA kit
Specificity This kit recognizes Human SPARCL1 in samples. No significant cross-reactivity or interference between Human SPARCL1 and analogues was observed.
Reproducibility Both intra-CV and inter-CV are

Information sourced from Uniprot.org


Item Specifications Storage
Micro CLIA Plate (Dismountable) 96T: 8 wells ×12 strips strips -20℃, 6 months
Reference Standard 96T: 2 vials 48T: 1 vial -20℃, 6 months
Concentrated Biotinylated Detection Ab (100×) 96T: 1 vial, 120 μL 60 μL -20℃, 6 months
Concentrated HRP Conjugate (100×) 96T: 1 vial, 120 μL 60 μL -20℃ (Protect from light), 6 months
Reference Standard & Sample Diluent 1 vial, 20 mL 2-8°C, 6 months
Biotinylated Detection Ab Diluent 1 vial, 14 mL 2-8°C, 6 months
HRP Conjugate Diluent 1 vial, 14 mL 2-8°C, 6 months
Concentrated Wash Buffer (25×) 1 vial, 30 mL 2-8°C, 6 months
Substrate Reagent A 1 vial, 5 mL 2-8℃ (Protect from light)
Substrate Reagent B 1 vial, 5 mL 2-8℃ (Protect from light)
Plate Sealer 5 pieces
Manual 1 copy
Certificate of Analysis 1 copy

Sample Type Range (%) Average Recovery (%)
Serum(n=8) 98-115 106
EDTA plasma(n=8) 88-102 95
Cell culture media(n=8) 92-106 99

Intra-assay Precision Intra-assay Precision Intra-assay Precision Inter-assay Precision Inter-assay Precision Inter-assay Precision
Sample 1.00 2.00 3.00 1.00 2.00 3.00
n 20.00 20.00 20.00 20.00 20.00 20.00
Mean (pg/mL) 197.73 553.13 1740.74 182.80 540.30 1831.46
Standard deviation 21.08 60.01 105.31 18.72 59.32 149.81
CV (%) 10.66 10.85 6.05 10.24 10.98 8.18
12 months for antibodies. 6 months for ELISA Kits. Please see website T&Cs for further guidance