Applications: |
CLIA |
Reactivity: |
Human |
Note: |
FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS. |
Sensitivity: |
37.5pg/mL |
Detection Limit: |
62.5~4000pg/mL |
Short Description: |
This human PPIB kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes. |
Storage Instruction: |
If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual. |
Assay Time: |
3.5h |
Detection: |
Chemiluminescence |
Gene Symbol: |
PPIB |
Gene ID: |
5479 |
Uniprot ID: |
PPIB_HUMAN |
Specificity: |
This kit recognizes Human CYPB in samples. No significant cross-reactivity or interference between Human CYPB and analogues was observed. |
Sample Type: |
Serum, plasma and other biological fluids |
Function | PPIase that catalyzes the cis-trans isomerization of proline imidic peptide bonds in oligopeptides and may therefore assist protein folding. |
Protein Name | Peptidyl-Prolyl Cis-Trans Isomerase BPpiase BCyp-S1Cyclophilin BRotamase BS-CyclophilinScylp |
Database Links | Reactome: R-HSA-1650814Reactome: R-HSA-9692916 |
Cellular Localisation | Virion(Microbial Infection)Endoplasmic Reticulum LumenMelanosomeIdentified By Mass Spectrometry In Melanosome Fractions From Stage I To Stage Iv |
Alternative CLIA Names | Peptidyl-Prolyl Cis-Trans Isomerase B CLIA kitPpiase B CLIA kitCyp-S1 CLIA kitCyclophilin B CLIA kitRotamase B CLIA kitS-Cyclophilin CLIA kitScylp CLIA kitPPIB CLIA kitCYPB CLIA kit |
Specificity | This kit recognizes Human CYPB in samples. No significant cross-reactivity or interference between Human CYPB and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
Item | Specifications | Storage |
Micro CLIA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
Substrate Reagent A | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Substrate Reagent B | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Plate Sealer | 5 pieces | |
Manual | 1 copy | |
Certificate of Analysis | 1 copy | |
Sample Type | Range (%) | Average Recovery (%) |
Serum(n=8) | 92-106 | 99 |
EDTA plasma(n=8) | 84-97 | 91 |
Cell culture media(n=8) | 96-111 | 102 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision |
Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
Mean (pg/mL) | 200.60 | 418.99 | 1835.95 | 186.21 | 405.29 | 1704.59 |
Standard deviation | 22.67 | 41.27 | 144.31 | 16.57 | 48.43 | 114.04 |
CV (%) | 11.30 | 9.85 | 7.86 | 8.90 | 11.95 | 6.69 |
12 months for antibodies. 6 months for ELISA Kits. Please see website T&Cs for further guidance