Applications: |
CLIA |
Reactivity: |
Human |
Note: |
FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS. |
Sensitivity: |
0.19ng/mL |
Detection Limit: |
0.31~20ng/mL |
Short Description: |
This human GP5 kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes. |
Storage Instruction: |
If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual. |
Assay Time: |
3.5h |
Detection: |
Chemiluminescence |
Gene Symbol: |
GP5 |
Gene ID: |
2814 |
Uniprot ID: |
GPV_HUMAN |
Specificity: |
This kit recognizes Human GP5 in samples. No significant cross-reactivity or interference between Human GP5 and analogues was observed. |
Sample Type: |
Serum, plasma and other biological fluids |
Tissue Specificity | Platelets and megakaryocytes. |
Post Translational Modifications | The N-terminus is blocked. |
Function | The GPIb-V-IX complex functions as the vWF receptor and mediates vWF-dependent platelet adhesion to blood vessels. The adhesion of platelets to injured vascular surfaces in the arterial circulation is a critical initiating event in hemostasis. |
Protein Name | Platelet Glycoprotein VGpvGlycoprotein 5Cd Antigen Cd42d |
Database Links | Reactome: R-HSA-140837Reactome: R-HSA-430116Reactome: R-HSA-75892Reactome: R-HSA-76009Reactome: R-HSA-9673221 |
Cellular Localisation | MembraneSingle-Pass Type I Membrane Protein |
Alternative CLIA Names | Platelet Glycoprotein V CLIA kitGpv CLIA kitGlycoprotein 5 CLIA kitCd Antigen Cd42d CLIA kitGP5 CLIA kit |
Specificity | This kit recognizes Human GP5 in samples. No significant cross-reactivity or interference between Human GP5 and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
Item | Specifications | Storage |
Micro CLIA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
Substrate Reagent A | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Substrate Reagent B | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Plate Sealer | 5 pieces | |
Manual | 1 copy | |
Certificate of Analysis | 1 copy | |
Sample Type | Range (%) | Average Recovery (%) |
Serum(n=8) | 83-95 | 90 |
EDTA plasma(n=8) | 96-110 | 104 |
Cell culture media(n=8) | 85-98 | 90 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision |
Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
Mean (ng/mL) | 0.98 | 1.68 | 9.11 | 1.07 | 1.79 | 8.62 |
Standard deviation | 0.08 | 0.15 | 0.82 | 0.13 | 0.18 | 0.66 |
CV (%) | 8.16 | 8.93 | 9.00 | 12.15 | 10.06 | 7.66 |
12 months for antibodies. 6 months for ELISA Kits. Please see website T&Cs for further guidance