• Human CLDN4 (Claudin-4) CLIA Kit (STJC0000417)

Human CLDN4 (Claudin-4) CLIA Kit (STJC0000417)

SKU:
STJC0000417

Shipping:
Free Shipping
Current Stock:
Applications: CLIA
Reactivity: Human
Note: FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS.
Sensitivity: 18.75pg/mL
Detection Limit: 31.25~2000pg/mL
Short Description: This human CLDN4 kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes.
Storage Instruction: If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual.
Assay Time: 3.5h
Detection: Chemiluminescence
Gene Symbol: CLDN4
Gene ID: 1364
Uniprot ID: CLD4_HUMAN
Specificity: This kit recognizes Human CLDN4 in samples. No significant cross-reactivity or interference between Human CLDN4 and analogues was observed.
Sample Type: Serum, plasma and other biological fluids
Post Translational Modifications Phosphorylated. Phosphorylation by EPHA2 is stimulated by EFNA1 and alters interaction with TJP1.
Function Channel-forming tight junction protein that mediates paracellular chloride transport in the kidney. Plays a critical role in the paracellular reabsorption of filtered chloride in the kidney collecting ducts. Claudins play a major role in tight junction-specific obliteration of the intercellular space, through calcium-independent cell-adhesion activity.
Protein Name Claudin-4
Clostridium Perfringens Enterotoxin Receptor
Cpe-R
Cpe-Receptor
Williams-Beuren Syndrome Chromosomal Region 8 Protein
Database Links Reactome: R-HSA-420029
Cellular Localisation Cell Junction
Tight Junction
Cell Membrane
Multi-Pass Membrane Protein
Cldn4 Is Required For Tight Junction Localization In The Kidney
Alternative CLIA Names Claudin-4 CLIA kit
Clostridium Perfringens Enterotoxin Receptor CLIA kit
Cpe-R CLIA kit
Cpe-Receptor CLIA kit
Williams-Beuren Syndrome Chromosomal Region 8 Protein CLIA kit
CLDN4 CLIA kit
CPER CLIA kit
CPETR1 CLIA kit
WBSCR8 CLIA kit
Specificity This kit recognizes Human CLDN4 in samples. No significant cross-reactivity or interference between Human CLDN4 and analogues was observed.
Reproducibility Both intra-CV and inter-CV are

Information sourced from Uniprot.org


Item Specifications Storage
Micro CLIA Plate (Dismountable) 96T: 8 wells ×12 strips strips -20℃, 6 months
Reference Standard 96T: 2 vials 48T: 1 vial -20℃, 6 months
Concentrated Biotinylated Detection Ab (100×) 96T: 1 vial, 120 μL 60 μL -20℃, 6 months
Concentrated HRP Conjugate (100×) 96T: 1 vial, 120 μL 60 μL -20℃ (Protect from light), 6 months
Reference Standard & Sample Diluent 1 vial, 20 mL 2-8°C, 6 months
Biotinylated Detection Ab Diluent 1 vial, 14 mL 2-8°C, 6 months
HRP Conjugate Diluent 1 vial, 14 mL 2-8°C, 6 months
Concentrated Wash Buffer (25×) 1 vial, 30 mL 2-8°C, 6 months
Substrate Reagent A 1 vial, 5 mL 2-8℃ (Protect from light)
Substrate Reagent B 1 vial, 5 mL 2-8℃ (Protect from light)
Plate Sealer 5 pieces
Manual 1 copy
Certificate of Analysis 1 copy

Sample Type Range (%) Average Recovery (%)
Serum(n=8) 93-105 100
EDTA plasma(n=8) 98-110 104
Cell culture media(n=8) 85-99 92

Intra-assay Precision Intra-assay Precision Intra-assay Precision Inter-assay Precision Inter-assay Precision Inter-assay Precision
Sample 1.00 2.00 3.00 1.00 2.00 3.00
n 20.00 20.00 20.00 20.00 20.00 20.00
Mean (pg/mL) 104.91 257.13 941.04 102.07 258.00 1013.35
Standard deviation 11.98 26.69 89.59 8.52 25.39 89.99
CV (%) 11.42 10.38 9.52 8.35 9.84 8.88
12 months for antibodies. 6 months for ELISA Kits. Please see website T&Cs for further guidance