Applications: |
CLIA |
Reactivity: |
Human |
Note: |
FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS. |
Sensitivity: |
4.69pg/mL |
Detection Limit: |
7.81~500pg/mL |
Short Description: |
This human AGRP kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes. |
Storage Instruction: |
If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual. |
Assay Time: |
3.5h |
Detection: |
Chemiluminescence |
Gene Symbol: |
AGRP |
Gene ID: |
181 |
Uniprot ID: |
AGRP_HUMAN |
Specificity: |
This kit recognizes Human AgRP in samples. No significant cross-reactivity or interference between Human AgRP and analogues was observed. |
Sample Type: |
Serum, plasma and other biological fluids |
Tissue Specificity | Expressed primarily in the adrenal gland, subthalamic nucleus, and hypothalamus, with a lower level of expression occurring in testis, lung, and kidney. |
Function | Plays a role in weight homeostasis. Involved in the control of feeding behavior through the central melanocortin system. Acts as alpha melanocyte-stimulating hormone antagonist by inhibiting cAMP production mediated by stimulation of melanocortin receptors within the hypothalamus and adrenal gland. Has very low activity with MC5R. Is an inverse agonist for MC3R and MC4R being able to suppress their constitutive activity. It promotes MC3R and MC4R endocytosis in an arrestin-dependent manner. |
Protein Name | Agouti-Related Protein |
Database Links | Reactome: R-HSA-9615017 |
Cellular Localisation | SecretedGolgi Apparatus Lumen |
Alternative CLIA Names | Agouti-Related Protein CLIA kitAGRP CLIA kitAGRT CLIA kitART CLIA kit |
Specificity | This kit recognizes Human AgRP in samples. No significant cross-reactivity or interference between Human AgRP and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
Item | Specifications | Storage |
Micro CLIA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
Substrate Reagent A | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Substrate Reagent B | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Plate Sealer | 5 pieces | |
Manual | 1 copy | |
Certificate of Analysis | 1 copy | |
Sample Type | Range (%) | Average Recovery (%) |
Serum(n=8) | 85-96 | 90 |
EDTA plasma(n=8) | 95-113 | 103 |
Cell culture media(n=8) | 91-105 | 99 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision |
Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
Mean (pg/mL) | 25.81 | 80.27 | 188.20 | 27.50 | 74.73 | 172.76 |
Standard deviation | 3.33 | 8.70 | 17.52 | 3.55 | 6.72 | 14.08 |
CV (%) | 12.90 | 10.84 | 9.31 | 12.91 | 8.99 | 8.15 |
12 months for antibodies. 6 months for ELISA Kits. Please see website T&Cs for further guidance