Applications: |
CLIA |
Reactivity: |
Rat |
Note: |
FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS. |
Sensitivity: |
0.19ng/mL |
Detection Limit: |
0.31~20ng/mL |
Short Description: |
This rat S100A11 kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes. |
Storage Instruction: |
If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual. |
Assay Time: |
3.5h |
Detection: |
Chemiluminescence |
Gene Symbol: |
S100a11 |
Gene ID: |
445415 |
Uniprot ID: |
S10AB_RAT |
Specificity: |
This kit recognizes Rat S100A11 in samples. No significant cross-reactivity or interference between Rat S100A11 and analogues was observed. |
Sample Type: |
Serum, plasma and other biological fluids |
Post Translational Modifications | Phosphorylation at Thr-5 significantly suppresses homodimerization and promotes association with NCL/nucleolin which induces nuclear translocation. |
Function | Facilitates the differentiation and the cornification of keratinocytes. |
Protein Name | Protein S100-A11CalgizzarinS100 Calcium-Binding Protein A11 |
Database Links | Reactome: R-RNO-6798695 |
Cellular Localisation | CytoplasmNucleus |
Alternative CLIA Names | Protein S100-A11 CLIA kitCalgizzarin CLIA kitS100 Calcium-Binding Protein A11 CLIA kitS100a11 CLIA kit |
Specificity | This kit recognizes Rat S100A11 in samples. No significant cross-reactivity or interference between Rat S100A11 and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
Item | Specifications | Storage |
Micro CLIA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
Substrate Reagent A | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Substrate Reagent B | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Plate Sealer | 5 pieces | |
Manual | 1 copy | |
Certificate of Analysis | 1 copy | |
Sample Type | Range (%) | Average Recovery (%) |
Serum(n=8) | 99-114 | 105 |
EDTA plasma(n=8) | 93-108 | 100 |
Cell culture media(n=8) | 90-105 | 96 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision |
Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
Mean (ng/mL) | 1.11 | 2.11 | 6.84 | 1.18 | 1.98 | 7.41 |
Standard deviation | 0.12 | 0.22 | 0.42 | 0.15 | 0.20 | 0.45 |
CV (%) | 10.81 | 10.43 | 6.14 | 12.71 | 10.10 | 6.07 |
12 months for antibodies. 6 months for ELISA Kits. Please see website T&Cs for further guidance