Applications: |
CLIA |
Reactivity: |
Rat |
Note: |
FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS. |
Sensitivity: |
1.88pg/mL |
Detection Limit: |
3.13~200pg/mL |
Short Description: |
This rat INHbC kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes. |
Storage Instruction: |
If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual. |
Assay Time: |
3.5h |
Detection: |
Chemiluminescence |
Gene Symbol: |
Inhbc |
Gene ID: |
64549 |
Uniprot ID: |
INHBC_RAT |
Specificity: |
This kit recognizes Rat INHbC in samples. No significant cross-reactivity or interference between Rat INHbC and analogues was observed. |
Sample Type: |
Serum, plasma and other biological fluids |
Function | Inhibins and activins inhibit and activate, respectively, the secretion of follitropin by the pituitary gland. Inhibins/activins are involved in regulating a number of diverse functions such as hypothalamic and pituitary hormone secretion, gonadal hormone secretion, germ cell development and maturation, erythroid differentiation, insulin secretion, nerve cell survival, embryonic axial development or bone growth, depending on their subunit composition. Inhibins appear to oppose the functions of activins. |
Protein Name | Inhibin Beta C ChainActivin Beta-C Chain |
Database Links | Reactome: R-RNO-209822 |
Cellular Localisation | Secreted |
Alternative CLIA Names | Inhibin Beta C Chain CLIA kitActivin Beta-C Chain CLIA kitInhbc CLIA kit |
Specificity | This kit recognizes Rat INHbC in samples. No significant cross-reactivity or interference between Rat INHbC and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
Item | Specifications | Storage |
Micro CLIA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
Substrate Reagent A | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Substrate Reagent B | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Plate Sealer | 5 pieces | |
Manual | 1 copy | |
Certificate of Analysis | 1 copy | |
Sample Type | Range (%) | Average Recovery (%) |
Serum(n=8) | 100-114 | 108 |
EDTA plasma(n=8) | 97-114 | 104 |
Cell culture media(n=8) | 94-107 | 101 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision |
Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
Mean (pg/mL) | 10.67 | 21.53 | 90.20 | 10.99 | 23.65 | 94.54 |
Standard deviation | 1.03 | 2.32 | 8.17 | 1.22 | 2.75 | 7.81 |
CV (%) | 9.65 | 10.78 | 9.06 | 11.10 | 11.63 | 8.26 |
12 months for antibodies. 6 months for ELISA Kits. Please see website T&Cs for further guidance