Applications: |
CLIA |
Reactivity: |
Rat |
Note: |
FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS. |
Sensitivity: |
3.75ng/mL |
Detection Limit: |
6.25~400ng/mL |
Short Description: |
This rat IgM kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes. |
Storage Instruction: |
If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual. |
Assay Time: |
3.5h |
Detection: |
Chemiluminescence |
Specificity: |
This kit recognizes Rat IgM in samples. No significant cross-reactivity or interference between Rat IgM and analogues was observed. |
Sample Type: |
Serum, plasma and other biological fluids |
Specificity | This kit recognizes Rat IgM in samples. No significant cross-reactivity or interference between Rat IgM and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
Item | Specifications | Storage |
Micro CLIA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
Substrate Reagent A | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Substrate Reagent B | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Plate Sealer | 5 pieces | |
Manual | 1 copy | |
Certificate of Analysis | 1 copy | |
Sample Type | Range (%) | Average Recovery (%) |
Serum(n=8) | 98-113 | 104 |
EDTA plasma(n=8) | 96-113 | 104 |
Cell culture media(n=8) | 86-100 | 91 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision |
Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
Mean (ng/mL) | 21.64 | 41.04 | 154.86 | 21.32 | 45.02 | 147.52 |
Standard deviation | 2.08 | 3.67 | 11.72 | 2.53 | 3.71 | 9.49 |
CV (%) | 9.61 | 8.94 | 7.57 | 11.87 | 8.24 | 6.43 |
12 months for antibodies. 6 months for ELISA Kits. Please see website T&Cs for further guidance