Applications: |
CLIA |
Reactivity: |
Mouse |
Note: |
FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS. |
Sensitivity: |
37.5pg/mL |
Detection Limit: |
62.5~4000pg/mL |
Short Description: |
This mouse ICAM-1 kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes. |
Storage Instruction: |
If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual. |
Assay Time: |
3.5h |
Detection: |
Chemiluminescence |
Gene Symbol: |
Icam1 |
Gene ID: |
15894 |
Uniprot ID: |
ICAM1_MOUSE |
Specificity: |
This kit recognizes Mouse ICAM-1 in samples. No significant cross-reactivity or interference between Mouse ICAM-1 and analogues was observed. |
Sample Type: |
Serum, plasma and other biological fluids |
Tissue Specificity | Expressed at low level on a subpopulation of lymphocytes, macrophages, and endothelial cells, but is strongly induced on these cells, and on fibroblasts and epithelial cells. |
Post Translational Modifications | Monoubiquitinated, which is promoted by MARCH9 and leads to endocytosis. |
Function | ICAM proteins are ligands for the leukocyte adhesion protein LFA-1 (integrin alpha-L/beta-2). During leukocyte trans-endothelial migration, ICAM1 engagement promotes the assembly of endothelial apical cups through ARHGEF26/SGEF and RHOG activation. |
Protein Name | Intercellular Adhesion Molecule 1Icam-1Mala-2Myd10Cd Antigen Cd54 |
Database Links | Reactome: R-MMU-198933Reactome: -MMU-216083 |
Cellular Localisation | MembraneSingle-Pass Type I Membrane Protein |
Alternative CLIA Names | Intercellular Adhesion Molecule 1 CLIA kitIcam-1 CLIA kitMala-2 CLIA kitMyd10 CLIA kitCd Antigen Cd54 CLIA kitIcam1 CLIA kitIcam-1 CLIA kit |
Specificity | This kit recognizes Mouse ICAM-1 in samples. No significant cross-reactivity or interference between Mouse ICAM-1 and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
Item | Specifications | Storage |
Micro CLIA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
Substrate Reagent A | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Substrate Reagent B | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Plate Sealer | 5 pieces | |
Manual | 1 copy | |
Certificate of Analysis | 1 copy | |
Sample Type | Range (%) | Average Recovery (%) |
Serum(n=8) | 94-106 | 99 |
EDTA plasma(n=8) | 95-108 | 103 |
Cell culture media(n=8) | 92-105 | 98 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision |
Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
Mean (pg/mL) | 191.99 | 578.30 | 1886.02 | 207.53 | 546.87 | 1712.80 |
Standard deviation | 16.28 | 42.39 | 146.54 | 24.76 | 57.31 | 105.51 |
CV (%) | 8.48 | 7.33 | 7.77 | 11.93 | 10.48 | 6.16 |
12 months for antibodies. 6 months for ELISA Kits. Please see website T&Cs for further guidance