• Mouse FABP5 (Fatty Acid Binding Protein 5, Epidermal) CLIA Kit (STJC0001214)

Mouse FABP5 (Fatty Acid Binding Protein 5, Epidermal) CLIA Kit (STJC0001214)

SKU:
STJC0001214

Shipping:
Free Shipping
Current Stock:
Applications: CLIA
Reactivity: Mouse
Note: FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS.
Sensitivity: 18.75pg/mL
Detection Limit: 31.25~2000pg/mL
Short Description: This mouse FABP5 kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes.
Storage Instruction: If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual.
Assay Time: 3.5h
Detection: Chemiluminescence
Specificity: This kit recognizes Mouse FABP5 in samples. No significant cross-reactivity or interference between Mouse FABP5 and analogues was observed.
Sample Type: Serum, plasma and other biological fluids
Specificity This kit recognizes Mouse FABP5 in samples. No significant cross-reactivity or interference between Mouse FABP5 and analogues was observed.
Reproducibility Both intra-CV and inter-CV are

Information sourced from Uniprot.org


Item Specifications Storage
Micro CLIA Plate (Dismountable) 96T: 8 wells ×12 strips strips -20℃, 6 months
Reference Standard 96T: 2 vials 48T: 1 vial -20℃, 6 months
Concentrated Biotinylated Detection Ab (100×) 96T: 1 vial, 120 μL 60 μL -20℃, 6 months
Concentrated HRP Conjugate (100×) 96T: 1 vial, 120 μL 60 μL -20℃ (Protect from light), 6 months
Reference Standard & Sample Diluent 1 vial, 20 mL 2-8°C, 6 months
Biotinylated Detection Ab Diluent 1 vial, 14 mL 2-8°C, 6 months
HRP Conjugate Diluent 1 vial, 14 mL 2-8°C, 6 months
Concentrated Wash Buffer (25×) 1 vial, 30 mL 2-8°C, 6 months
Substrate Reagent A 1 vial, 5 mL 2-8℃ (Protect from light)
Substrate Reagent B 1 vial, 5 mL 2-8℃ (Protect from light)
Plate Sealer 5 pieces
Manual 1 copy
Certificate of Analysis 1 copy

Sample Type Range (%) Average Recovery (%)
Serum(n=8) 91-106 97
EDTA plasma(n=8) 95-106 100
Cell culture media(n=8) 91-103 98

Intra-assay Precision Intra-assay Precision Intra-assay Precision Inter-assay Precision Inter-assay Precision Inter-assay Precision
Sample 1.00 2.00 3.00 1.00 2.00 3.00
n 20.00 20.00 20.00 20.00 20.00 20.00
Mean (pg/mL) 109.61 288.63 734.19 108.87 300.01 669.32
Standard deviation 14.22 21.36 71.88 13.50 27.60 62.72
CV (%) 12.97 7.40 9.79 12.40 9.20 9.37
12 months for antibodies. 6 months for ELISA Kits. Please see website T&Cs for further guidance