Applications: |
CLIA |
Reactivity: |
Human |
Note: |
FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS. |
Sensitivity: |
9.38pg/mL |
Detection Limit: |
15.63~1000pg/mL |
Short Description: |
This human DEFB1 kit is a highly sensitive in-vitro chemiluminescent immunoassay for the measurement of trace amounts of analytes. |
Storage Instruction: |
If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual. |
Assay Time: |
3.5h |
Detection: |
Chemiluminescence |
Gene Symbol: |
DEFB1 |
Gene ID: |
1672 |
Uniprot ID: |
DEFB1_HUMAN |
Specificity: |
This kit recognizes Human DEFB1 in samples. No significant cross-reactivity or interference between Human DEFB1 and analogues was observed. |
Sample Type: |
Serum, plasma and other biological fluids |
Tissue Specificity | Blood plasma. Sperm. Highly expressed in the lower head and midpiece of sperm. Significantly reduced levels found in the sperms of asthenozoospermia and leukocytospermia patients (at protein level). |
Function | Has bactericidal activity. May act as a ligand for C-C chemokine receptor CCR6. Positively regulates the sperm motility and bactericidal activity in a CCR6-dependent manner. Binds to CCR6 and triggers Ca2+ mobilization in the sperm which is important for its motility. |
Protein Name | Beta-Defensin 1Bd-1Hbd-1Defensin - Beta 1 |
Database Links | Reactome: R-HSA-1461957Reactome: R-HSA-1461973 |
Cellular Localisation | SecretedMembraneAssociates With Tumor Cell Membrane-Derived Microvesicles |
Alternative CLIA Names | Beta-Defensin 1 CLIA kitBd-1 CLIA kitHbd-1 CLIA kitDefensin - Beta 1 CLIA kitDEFB1 CLIA kitBD1 CLIA kitHBD1 CLIA kit |
Specificity | This kit recognizes Human DEFB1 in samples. No significant cross-reactivity or interference between Human DEFB1 and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
Item | Specifications | Storage |
Micro CLIA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
Substrate Reagent A | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Substrate Reagent B | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Plate Sealer | 5 pieces | |
Manual | 1 copy | |
Certificate of Analysis | 1 copy | |
Sample Type | Range (%) | Average Recovery (%) |
Serum(n=8) | 94-107 | 99 |
EDTA plasma(n=8) | 91-105 | 98 |
Cell culture media(n=8) | 87-101 | 94 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision |
Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
Mean (pg/mL) | 50.48 | 150.66 | 477.74 | 46.71 | 140.15 | 469.68 |
Standard deviation | 5.14 | 13.73 | 42.09 | 5.06 | 10.75 | 43.12 |
CV (%) | 10.18 | 9.11 | 8.81 | 10.83 | 7.67 | 9.18 |
12 months for antibodies. 6 months for ELISA Kits. Please see website T&Cs for further guidance