Human CEACAM5 (Carcinoembryonic antigen-related cell adhesion molecule 5) Sandwich ELISA (STJE0004059)

SKU:
STJE0004059

Shipping:
Free Shipping
Current Stock:
Applications: ELISA
Reactivity: Human
Note: FOR SCIENTIFIC EDUCATIONAL RESEARCH USE ONLY (RUO). MUST NOT BE USED IN DIAGNOSTIC OR OTHER MEDICAL APPLICATIONS.
Sensitivity: 0.1ng/mL
Detection Limit: 0.16~10ng/mL
Short Description: This CEACAM5 Sandwich ELISA is an in-vitro enzyme-linked immunosorbent assay for the measurement of samples in human cell culture supernatant, serum and plasma (EDTA, citrate, heparin).
Storage Instruction: If unopened the kit may be stored at 2-8°C for up to 1 month. If the kit will not be used within 1 month, store the components separately, according to the component table in the manual.
Assay Time: 3.5h
Detection: Colormetric
Gene Symbol: CEACAM5
Gene ID: 1048
Uniprot ID: CEAM5_HUMAN
Specificity: This kit recognizes Human CEA in samples. No significant cross-reactivity or interference between Human CEA and analogues was observed.
Sample Type: Serum, plasma and other biological fluids
Post Translational Modifications Complex immunoreactive glycoprotein with a MW of 180 kDa comprising 60% carbohydrate.
Function Cell surface glycoprotein that plays a role in cell adhesion, intracellular signaling and tumor progression. Mediates homophilic and heterophilic cell adhesion with other carcinoembryonic antigen-related cell adhesion molecules, such as CEACAM6. Plays a role as an oncogene by promoting tumor progression.induces resistance to anoikis of colorectal carcinoma cells. (Microbial infection) Receptor for E.coli Dr adhesins. Binding of E.coli Dr adhesins leads to dissociation of the homodimer.
Protein Name Carcinoembryonic Antigen-Related Cell Adhesion Molecule 5
Carcinoembryonic Antigen
Cea
Meconium Antigen 100
Cd Antigen Cd66e
Database Links Reactome: R-HSA-163125
Reactome: R-HSA-202733
Cellular Localisation Cell Membrane
Lipid-Anchor
Gpi-Anchor
Apical Cell Membrane
Cell Surface
Localized To The Apical Glycocalyx Surface
Alternative ELISA Names Carcinoembryonic Antigen-Related Cell Adhesion Molecule 5 ELISA kit
Carcinoembryonic Antigen ELISA kit
Cea ELISA kit
Meconium Antigen 100 ELISA kit
Cd Antigen Cd66e ELISA kit
CEACAM5 ELISA kit
CEA ELISA kit
Specificity This kit recognizes Human CEA in samples. No significant cross-reactivity or interference between Human CEA and analogues was observed.
Reproducibility Both intra-CV and inter-CV are

Information sourced from Uniprot.org


Item Specifications Storage
Micro ELISA Plate (Dismountable) 96T: 8 wells ×12 strips strips -20℃, 6 months
Reference Standard 96T: 2 vials 48T: 1 vial -20℃, 6 months
Concentrated Biotinylated Detection Ab (100×) 96T: 1 vial, 120 μL 60 μL -20℃, 6 months
Concentrated HRP Conjugate (100×) 96T: 1 vial, 120 μL 60 μL -20℃ (Protect from light), 6 months
Reference Standard & Sample Diluent 1 vial, 20 mL 2-8°C, 6 months
Biotinylated Detection Ab Diluent 1 vial, 14 mL 2-8°C, 6 months
HRP Conjugate Diluent 1 vial, 14 mL 2-8°C, 6 months
Concentrated Wash Buffer (25×) 1 vial, 30 mL 2-8°C, 6 months
Substrate Reagent 1 vial, 10 mL 2-8℃ (Protect from light)
Stop Solution 1 vial, 10 mL 2-8°C
Plate Sealer 5 pieces
Manual 1 copy
Certificate of Analysis 1 copy

Sample Type Range (%) Average Recovery (%)
Serum(n=8) 93-104 96
EDTA plasma (n=8) 93-103 96
Cell culture media (n=8) 92-101 95

Intra-assay Precision Intra-assay Precision Intra-assay Precision Inter-assay Precision Inter-assay Precision Inter-assay Precision
Sample 1.00 2.00 3.00 1.00 2.00 3.00
n 20.00 20.00 20.00 20.00 20.00 20.00
Mean (ng/mL) 0.49 1.20 4.32 0.50 1.04 4.59
Standard deviation 0.03 0.06 0.18 0.03 0.05 0.19
CV (%) 6.12 5.00 4.17 6.00 4.81 4.14
12 months for antibodies. 6 months for ELISA Kits. Please see website T&Cs for further guidance