Post Translational Modifications | The N-terminal processed forms of platelet factor 4 variant seems to be produced by proteolytic cleavage. The most abundant form is Platelet factor 4 variant(5-74). |
Function | Inhibitor of angiogenesis. Inhibitor of endothelial cell chemotaxis (in vitro). |
Protein Name | Platelet Factor 4 VariantC-X-C Motif Chemokine 4 VariantCxcl4l1Pf4altPf4var1 Cleaved Into - Platelet Factor 4 Variant(4-74 - Platelet Factor 4 Variant(5-74 - Platelet Factor 4 Variant(6-74 |
Database Links | Reactome: R-HSA-140875Reactome: R-HSA-202733 |
Cellular Localisation | Secreted |
Alternative CLIA Names | Platelet Factor 4 Variant CLIA kitC-X-C Motif Chemokine 4 Variant CLIA kitCxcl4l1 CLIA kitPf4alt CLIA kitPf4var1 Cleaved Into - Platelet Factor 4 Variant(4-74 - Platelet Factor 4 Variant(5-74 - Platelet Factor 4 Variant(6-74 CLIA kitPF4V1 CLIA kitCXCL4V1 CLIA kitSCYB4V1 CLIA kit |
Specificity | This kit recognizes Human PF4V1 in samples. No significant cross-reactivity or interference between Human PF4V1 and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are |
Information sourced from Uniprot.org
Item | Specifications | Storage |
Micro CLIA Plate (Dismountable) | 96T: 8 wells ×12 strips strips | -20℃, 6 months |
Reference Standard | 96T: 2 vials 48T: 1 vial | -20℃, 6 months |
Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃, 6 months |
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 60 μL | -20℃ (Protect from light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
HRP Conjugate Diluent | 1 vial, 14 mL | 2-8°C, 6 months |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | 2-8°C, 6 months |
Substrate Reagent A | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Substrate Reagent B | 1 vial, 5 mL | 2-8℃ (Protect from light) |
Plate Sealer | 5 pieces | |
Manual | 1 copy | |
Certificate of Analysis | 1 copy | |
Sample Type | Range (%) | Average Recovery (%) |
Serum(n=8) | 89-103 | 95 |
EDTA plasma(n=8) | 86-99 | 91 |
Cell culture media(n=8) | 92-106 | 98 |
| Intra-assay Precision | Intra-assay Precision | Intra-assay Precision | Inter-assay Precision | Inter-assay Precision | Inter-assay Precision |
Sample | 1.00 | 2.00 | 3.00 | 1.00 | 2.00 | 3.00 |
n | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 | 20.00 |
Mean (ng/mL) | 0.99 | 2.08 | 6.99 | 1.00 | 2.22 | 6.63 |
Standard deviation | 0.12 | 0.22 | 0.65 | 0.08 | 0.22 | 0.63 |
CV (%) | 12.12 | 10.58 | 9.30 | 8.00 | 9.91 | 9.50 |